SciTemp® Pre-Calibrated, Disposable Temperature Sensors
The pre-calibrated and disposable SciTemp is a cost effective temperature sensing solution for a wide array of liquid handling applications (e.g.: liquid chromatography, fermentaiton, disposable bag manifolds, filtration, etc.). A unique sensor ID and unit specific calibration data is pre-programmed into each unit's embedded ROM chip. The SciTemp monitor can display readouts from up to 2 temperature sensors simultaneously and displays T1, T2, and Differential Temperature.
| Material, Fluid Contact: |
Medical Grade Polysulfone - meets USP Class VI and FDA 21 CFR 177.1520.
All Wetted Materials are made of Animal-Free Compounds. Compatible with most sanitizing agents such as NaOH, Hypochlorite. The sensors can also be repeatedly autolaved.
|
| Sensor Type: |
Thermisor, Epoxy-Coated, 2252 Ohms |
| Temperature Range: |
-40 to +150 °C |
| Accuracy: |
+/- 0.10 °C in the 4.00 to 70.00 °C Range |
| Temperature Display Resolution: |
Two Decimal Places: 0.01 °C |
| Sensor Cable & Connectors |
Lockable & Waterproof |
| Sensor Microchip |
EPROM, Stored Sensor ID and Cal. Factor. |
Available Connector Types: |
Luer, 3/8" Barb, 1/2" Barb, 3/4" TC (Tri-Clover), and 1.0" TC (Tri-Clover) |
Compatible Tubing Sizes:
|
Luer: ID 0.03" to 0.31"
3/8" Barb: ID 0.31" to 0.38"
1/2" Barb: ID 0.50"
3/4" TC: Tubing with 3/4" TC
1.0" TC: Tubing with 1" TC Ladish
|
Maximum Flow Rate: |
Luer: 1 liter/min
3/8" Barb: 8 liters/min
1/2" Barb: 17 liters/min
3/4" TC: 31 liters/min
1.0" TC: 60 liters/min
|
Maximum Pressure:
|
Luer: 60 psi
3/8" Barb: 60 psi
1/2" Barb: 60 psi
3/4" TC: 90 psi
1.0" TC: 90 psi
|
Part Numbers: (Sold in packs of 5)
|
Luer: 080-799PS-5
3/8" Barb: 080-794PS-5
1/2" Barb: 080-795PS-5
3/4" TC: 080-796PS-5
1.0" TC: 808-797PS-5
|
*Patents Pending
© Copyright SciLog, Inc.
SciCon® Pre-Calibrated, Disposable Conductivity & Temperature Sensors
Bioprocessing with single-use disposable fluid paths is now possible.
The SciCon disposable, pre-calibrated conducitivity and temperature flow cells ship preprogramed with a unique sensor ID and unit specific calibration data. The SciCon monitor and sensor system is flexible enough for most liquid handling applications, including: Tangential Flow Filtration, Diafiltration, Ultrafiltration, Chromatography, Bag Manifolds, and Reverse Osmosis.
| Material, Fluid Contact: |
Natural PP, Gold, The Natural PP meets USP Class VI and FDA 21 CFR 177.1520 |
| Sensor Type: |
4-electrode conducitivity cell, Factory calibrated, Ready to use |
| Conductivity Range: |
1 uS/cm to 200 mS/cm |
| Accuracy: |
High Range: +/- 2.5 mS in the 10-200 mS Range
Low Range: +/- 3 µS in the 0-100 µS Range |
| Temperature Range: |
4 - 40° Celcius |
| Temperature Probe: |
Thermistor, Factory Calibrated |
| Temperature Accuracy: |
+/- 0.7° Celsius |
| Sensor Microchip: |
EPROM, stores Device ID, Cell Constant, Temp Offset & Factory Cal Data |
Available Connector Types: |
Luer, 3/8" Barb, 1/2" Barb, 3/4" TC (Tri-Clover), and 1.0" TC (Tri-Clover) |
Compatible Tubing Sizes:
|
Luer: ID 0.03" to 0.31"
3/8" Barb: ID 0.31" to 0.38"
1/2" Barb: ID 0.50"
3/4" TC: Tubing with 3/4" TC
1.0" TC: Tubing with 1" TC Ladish
|
Maximum Flow Rate: |
Luer: 2 liters/min
3/8" Barb: 8 liters/min
1/2" Barb: 17 liters/min
3/4" TC: 30.5 liters/min
1.0" TC: 60 liters/min
|
Maximum Pressure:
|
Luer: 65 psi
3/8" Barb: 50 psi
1/2" Barb: 50 psi
3/4" TC: 80 psi
1.0" TC: 100 psi
|
| Part Numbers:
|
Luer: 080-599
3/8" Barb: 080-594
1/2" Barb: 080-595
3/4" TC: 080-596
1.0" TC: 080-597PP (polypropylene),
1.0" TC: 808-597PS (polysulfone)
|
*Patents Pending
© Copyright SciLog, Inc.
SciPres® Pre-Calibrated, Disposable Pressure Sensors
SciPres is a cost effective, pre-calibrated, and disposable pressure sensing solution for a variety of liquid handling applications. Each pressure flow cell's internal chips are pre-programed with unique sensor ID and unit specific calibration data. The SciPres monitor can display readouts from up to 3 pressure sensors simultaneously and displays P1, P2, P3, and either Differential Pressure or Transmembrane Pressure.
| Material, Fluid Contact: |
Natural PP meet USP Class VI and FDA 21 CFR 177.1520. All wetted materials are made of animal-free compounds. Compatible with most sanitizing agents such as NaOH for flow cell sanitization. |
| Sensor Type: |
Medical Grade, Piezoresistive Sensing Element with on-chip temperature compensation |
| Sensor Isolation: |
Insoluble silicone dielectric gel isolates sensing element from process solution. The gel is a non-toxic, non-allergenic elastomer system that meets all USP XX Biological Class V requirements. |
| Pressure Range: |
0 - 60 psi: Luer, 0.375" Barb and 0.500" Barb Sensors.
0 - 90 psi: for 0.750" TC and 1.00" TC Ladish Sensors |
Pressure Accuracy: |
+/- 0.30 psi at 60 psi |
Pressure Resolution: |
0.01 psi Temperature Range: 0-60° C |
| Sensor Microchip: |
EPROM, Stored Sensor ID and Calibration Factor |
Sensor Connector Cables: |
Lockable & Waterproof |
| Sanitization: |
Can be sanitizied with NaOH, NaOCl, Formaldehyde, etc |
Available Connector Types: |
Luer, 3/8" Barb, 1/2" Barb, 3/4" TC, and 1.0" TC |
Compatible Tubing Sizes:
|
Luer: ID 0.03" to 0.31"
3/8" Barb: ID 0.31" to 0.38"
1/2" Barb: ID 0.50"
3/4" TC: Tubing with 3/4" TC
1.0" TC: Tubing with 1" TC Ladish
|
Maximum Flow Rate: |
Luer: 2 liters/min
3/8" Barb: 8 liters/min
1/2" Barb: 17 liters/min
3/4" TC: 30.5 liters/min
1.0" TC: 60 liters/min
|
Maximum Pressure:
|
Luer: 60 psi
3/8" Barb: 60 psi
1/2" Barb: 60 psi
3/4" TC: 100 psi
1.0" TC: 100 psi
|
| Part Numbers:
|
Luer: 080-693
3/8" Barb: 080-691
1/2" Barb: 080-692
3/4" TC: 080-698
1.0" TC: 080-697
|
*Patents Pending