| I/O Ports: DB 15 (Back Panel) |
1. RS-232 Serial port, DB15 connector, pin 2 (Transmit), pin 3 (Receive) & pin 5 (Ground)
2. 4-20 mA input signal, DB 15 connector, pin 6 (Ground) & pin 13, with selection switch (back panel) at 4-20 mA
3. 0-5VDC input signal, DB 15 connector, pin 6 (Ground) & pin 13, with selection switch (back panel) at 0-5VDC
4. Pump Speed output signal, DB 15 connector, pulsed output between pins 7 (Ground) & pin 9, pulse rate is proportional to pump speed, ~5940 pulses/sec at 99% pump speed, ~3000 pulses/sec at 50% pump speed.
5. Remote Start/Stop ( Remote ), closed contact closure (RUN), open contact closure (STOP) between pin 8 (Ground) & pin 15 of the DB 15 connector. This function is active when REMOTE has been selected from the front panel.
6. Remote CW/CCW Motor Direction ( Remote ), closed contact closure (CCW), open contact closure (CW) between pin 7 (Ground) & pin 14. This function is active when the REMOTE mode has been selected from the front panel.
7. Remote Start/Stop ( Local ), closed contact closure (RUN), open contact closure (STOP) between pin 8 (Ground) & pin 15 of the DB 15 connector. This function is active when the LOCAL mode has been selected from the front panel and the external contact closures (CC across pin 8 & 15) has been exercised, i.e. opening and closing the connected contact closures (CC). The front panel RUN/STOP key will be disabled at that point, however, the pump speed selection and the CW/CCW keys are still functional. The front panel RUN/STOP key function can be re-established by twice pressing the LOCAL/REMOTE key. NOTE: This operational mode should only be used when the ACCU pump is to be controlled (Stop/Start) by a SciLog Smart Pump, i.e. UltraTec or ChemTec.